8. Acoustic Environment

Purchase cart Previous page Return to chapter overview Next page

 

The acoustic environment must comply with the provisions of SANS 10182 with respect to maximum permissible ambient sound pressure levels (MPASPL).

 

8.1 Certification of Acoustic Environment
8.1.1 Audiometry must not be conducted unless the acoustic environment has been certified in accordance with the requirements of SANS 10182.
8.1.2 The certification must be conducted by a competent person.
8.1.3 The competent person must issue a certificate of compliance for that acoustic environment, which must contain the information required in the relevant parts of SANS 10182.
8.1.4 All acoustic environments must be certified before initial use and thereafter at intervals not exceeding 12-months: Provided that, should there be any change that may affect the ambient noise in the acoustic environment, the certification must be reissued immediately.
8.1.5 The certificate of compliance may only be issued once any deviations or defects observed have been corrected.
8.1.6 An audiometer must not be moved to an acoustic environment that is not certified.
8.1.7 No other medical examinations and tests must be conducted in the acoustic environment during the same time that an audiometric test is being conducted.